IS 17718 : 2022 Performance Testing of In-Vitro Diagnostics (IVD) Reagents/Kits - a-Amylase (CNP-G3), Liquid Stable Clinical Chemistry Reagents/Kits

ICS 11.100.10

MHD 19

New Standard from Last Update.

1. SCOPE

1.1 This standard describes the specifications and standard testing procedure, for determining the performance of In–vitro Diagnostics (IVD) clinical chemistry reagents/kits, used for the estimation of a-Amylase in human serum or plasma, based on the CNP-G3 method.

1.2 This test method is recommended for all forms of a-Amylase (CNP-G3), liquid stable clinical chemistry reagents/kits except those used in dry chemistry analyzers and closed fully automated chemistry analysers.