IS 2480 : Part 2 : 2024 General Purpose Glass Thermometers - Specification - Part 2 Enclosed-Scale Thermometers
1 SCOPE
This standard (Part 2) prescribes the requirements for enclosed-scale general purpose liquid-in-glass thermometers graduated for vertical total and partial immersion, covering the overall range - 100 ºC to 500 ºC.
2 REFERENCES
The standards given below contain provisions which through reference in this text, constitute provisions of and necessary adjuncts to this standard. At the time of publication, the editions indicated were valid. All standards are subject to revision and parties to agreements based on this standard are encouraged to investigate the possibility of applying the most recent edition of these standards:
IS No. | Title |
IS 321 : 1964 | Specification for absolute alcohol (first revision) |
IS 537 : 2011 | Toluene - Specification (second revision) |
IS 2627 : 1979 | Glossary of terms relating to liquid-in-glass thermometers (first revision) |
IS 4905 : 2015/ISO 24153 : 2009 | Random sampling and randomization procedures (first revision) |
IS 6274 : 1971 | Method of calibrating liquid-inglass thermometers |